Consulting
MWD Gamma Calibration,
Software and Training About Minette
Consulting
Sample
Calibration
Calibrator |
|
Calibrator API Value |
197 |
Calibrator cps for the First 5 Minutes |
193.2 |
Background cps |
21.2 |
Calibrator cps for the Last 5 minutes |
194.6 |
Average Calibrator cps |
193.9 |
Net Calibrator cps (Shielding Compensated) |
178.21 |
Probe Cal Ratio (API/cps) |
1.105 |
All the user has to do is choose the worksheet corresponding
to the proper calibrator (provided by Minette Consulting), and type in/copy
three numbers:
1)
Calibrator cps for the First 5
Minutes
2)
Background cps
3)
Calibrator cps for the Last 5
minutes
Weight: 70
lb
Length 12”
ID: 2.5”
OD: 10”
Features:
Directly
tied to the API test pits via the Minette Consulting Master Calibrator
API values
established to within 1 API (Typical Value 210 API).
Accommodates
probes up to 2.4” in diameter
Stainless
Steel Construction
Oilfield
Rugged and Reliable
Hinged for
Ease of Use